1. Sterilant residual-free system
2. Ergonomic design / Single-use cassette type sterilant
3. Thermal Printer
4. Table-top design
|
DESCRIPTION |
RENO – S20 |
|
|
TOTAL CYCLE TIME
|
Eco Cycle : 21min | |
| Advanced Cycle : 38min | ||
|
PROCESSING TEMPERATURE
|
Below 55°C | |
|
DIMENSIONS & WEIGHT |
Dimensions (H x W x Dmm) | 470 x 655 x 640 |
| Weight | 116Kg | |
|
STERILIZATION CHAMBER |
Configuration | Rectangular |
| Total Volume (H x W x Dmm) | 18L, 170 x260 x410 | |
| Usable Volume | 17L | |
| Material | Stainless Steel | |
|
STERILANT |
Liquid Chemical Agent | Hydrogen Peroxide |
| Concentration & Usage | 3.6ml (50%) | |
| 1Cycle / Cassette | ||
|
CONNECTION |
Electricity | 220V~, 단상, 60Hz |
|
PROGRAM |
Self-Test Mode | Yes |
| Printer Outputs | Thermal Printer | |
| Data Storage | Memory Card | |
|
DISPLAY |
Screen | 4″ TFT Touch Screen |
| Alarm | Audible Alarm | |
|
STERILIZES A WIDE RANGE OF STANDARD (CE) |
Single-Channel Stainless Steel Lumen | Eco Cycle: ≥ 1Ø x 500mm |
| Advanced Cycle: ≥ 1Ø x 1,000mm | ||
| Single-Channel Teflon Flexible Lumen | Eco Cycle: ≥ 1Ø x 1,500mm | |
| Advanced Cycle: ≥ 1Ø x 12,000mm | ||
| * Single-through Lumen | Advanced Cycle: ≥ 2Ø x 1,500mm | |
|
AIR INLET BACTERIA ETENTION FILTER |
HEFA filter | 99.97% Efficiency at 0.3 micron(0.3 x 10-6) |